EFFECTS OF SMOOTH, MEDIUM SMOOTH AND MEDIUM RECONSTRUCTION KERNELS ON IMAGE QUALITY IN THREE-PHASE CT OF LIVER
Keywords:
CT liver, image quality, medium kernel, medium smooth kernel, smooth kernelAbstract
Reconstruction kernel is one of the parameters that affects the computed tomography (CT) image quality. This study aimed to evaluate the effects of applying three different reconstruction kernels on image quality in 3-phased CT of the liver. A total of 63 CT liver images including normal liver (n = 43) and liver lesion (n = 20) were retrospectively reviewed. Smooth (B20f), medium smooth (B30f) and medium (B40f) reconstruction kernels were employed in the image reconstruction process. Mean attenuation, image noise, and signal-to-noise ratio (SNR) values from each kernel reconstruction were quantified and compared among those kernels using One Way Analysis of Variance (ANOVA) statistical analysis. Significant changes in image noise and SNR were observed in the normal liver (p < 0.001, respectively) following the application of those reconstruction kernels. However, no significant changes in mean attenuation, image noise, and SNR were demonstrated in the liver lesion (p > 0.05). Application of smooth (B20f), medium smooth (B30f), and medium (B40f) kernel reconstructions would significantly affect the image noise and SNR in the normal liver of CT images instead of liver lesions. Hence, proper selection of reconstruction kernel is important in CT images reconstruction to improve precision in diagnostic CT interpretation.
Downloads
Metrics
Published
How to Cite
Issue
Section
Any reproduction of figures, tables and illustrations must obtain written permission from the Chief Editor (wicki@ukm.edu.my). No part of the journal may be reproduced without the editor’s permission